ES601-US OPERATING MANUAL
Manual Testing/Chapter 5
#
Page 57
CALIBRATE
OFFSET:
The "
OFFSET:
" displayed to the right of this control is the resistance of the test
lead used in measuring point-to-point resistance. In a point-to-point test, the
OFFSET
calibration is subtracted from the measured resistance to obtain the
result which is displayed by ES601-US.
Calibrate the
OFFSET
when required, particularly if the test lead used for point-
to-point measurement is changed. Performing the calibration from System
Settings Menu #2 (see Section 3.3) records the
OFFSET
in non-volatile memory.
Performing the calibration from the Sequence Controls menu updates the
OFFSET
for the manual test, but does not store the calibration in the non-volatile
memory.
Calibrate the
OFFSET
according to this procedure:
1.
Connect the Kelvin Cable to the corresponding jacks of ES601-US.
2.
Connect the test lead which will be used for point-to-point
measurement to the "TO DUT CASE" jack of ES601-US.
3.
Connect the free ends of the Kelvin Cable and the test lead together.
Verify a solid mechanical contact exists between the two cables at the
far end.
4.
Press the
CALIBRATE OFFSET
button. The resistance will be
measured and displayed within one second.
5.
The result should be 0.05 ohms or lower for a typical test lead of 2-
meter length. For best results, avoid use of a poor-quality or excessively
long cable having a resistance exceeding 0.1 ohm. ES601-US will
produce an alarm and reject the calibration if the resistance exceeds 3
ohms.
6.
If the resistance seems unreasonably high, verify the connection
between the test lead and the Kelvin Cable. Also confirm the test lead
makes a solid contact when inserted into the "TO DUT CASE" jack.
Repeat the calibration as required until an acceptable result is obtained.
ESCAPE:
This button on the Sequence Controls menu aborts the test in progress and
returns the display to the MANUAL MODE menu.
NEXT:
Advances the test to the next step, which starts the resistance measurement.
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...