Figure 35
- Example setup for ECG monitor with 4 patient connections.
AAMI, IEC and VDE safety standards require patient leads or "applied parts" to be tested in related
Groups
.
For the AAMI-ES1 standard, there are only two Groups defined, organized by
electrical characteristics
:
Group 1 is pre-assigned to patient leads which are
electrically isolated
from AC supply ground (Type "BF");
Group 2 is pre-assigned to patient leads which are
not
isolated from ground (Type "B").
For IEC and VDE safety standards, Groups are organized by
medical function
: ECG, defib paddles,
blood pressure transducer, etc. ES601-US provides connections for up to ten patient leads via the AP jacks
located on the front panel, where the AP jacks can be organized in up to
five
functional Groups.
The assignment of patient leads to Groups is indicated on the PATIENT LEADS menu. Lead assignments
can be changed by pressing the
EDIT
button on the menu, which then displays the GROUPS editor:
The buttons appearing on the GROUPS editor are:
LEAD GROUP:
Selects the group of leads to be edited (1 or 2 for AAMI; 1, 2 3, 4 or 5 for IEC and
VDE).
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...