ES601-US OPERATING MANUAL
Manual Testing/Chapter 5
#
Page 72
If the leakage test is paused, the keypad buttons will be changed to:
ESCAPE:
Aborts the sequence in progress and return to the MANUAL MODE menu. No
leakage results are copied to the temporary data buffer.
REPEAT:
Repeats the measurement for the test conditions listed at the top of the display.
CONTINUE:
Allows the sequence to proceed to the next measurement.
If a value exceeding either of the posted limits is obtained, the sequence is automatically stopped at that
measurement and an audible alarm is produced by ES601-US. In this case, the display will offer the same
functions described above.
For IEC60601 and AAMI-ES1 standards, AC supply conditions will be tested in the following order,
according to test conditions initially specified by the user on the leakage setup menu:
1
Reverse polarity; Neutral in single-fault condition (
RP:OPEN N.
)
2
Reverse polarity; supply in normal condition (
RP:NORM.
) or with Ground in single-
fault condition (
RP:OPEN G.
)
3
Normal polarity; Neutral in single-fault condition (
NP:OPEN N.
)
4
Normal polarity; supply in normal condition (
NP:NORM.
) or with Ground in single-
fault condition (
NP:OPEN G.
)
At each configuration (1.)-(4.) of the AC supply, measurements of external and internal case leakage are
performed. Following these measurements, patient and lead-to-lead leakage will be measured if those tests
were enabled by the user on the setup menu.
When ES601-US changes the AC supply from neutral-open to normal condition, and if "STOP AFTER..."
of the Sequence Controls is "NO" (
Figure 34
), ES601-US will pause according to the "POWER-UP
DELAY" setting before proceeding to measure the leakage. This delay of 2 to 9 seconds allows the DUT to
startup up and stabilize before leakage readings are taken. Alternatively, if "STOP AFTER..." of the
Sequence Controls is "YES", ES601-US will change the AC supply from neutral-open to normal condition
and wait until you press the
CONTINUE
button:
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...