ES601-US OPERATING MANUAL
Setup/Chapter 3
#
Page 28
Figure 7
- System Settings Menu #3
3.4 System Settings Menu #3
The buttons provided on System Settings Menu #3 are:
SET CLK:
Allows setting the internal real time clock.
FORMAT:
Select date format for the internal real-time clock, as follows:
• USA (MMM DD YYYY)
• Europe (DD MMM YYYY)
• ISO (YYYY MMM DD)
EDIT ID:
ES601-US I.D. is set at time of manufacture to "MY ES601", and can be
edited using this utility which allows the user to input a unique I.D. to
identify ES601-US within the user's equipment management system. This
I.D. is meant to be a control number or similar reference, and should not
be confused with the unit serial number.
Pressing this button will activate the QWERTY touch keypad of ES601-
US, allowing the user to type in an I.D. Refer to Section 3.5 for more
information on operation and features of the QWERTY keypad.
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...