ES601-US OPERATING MANUAL
Setup/Chapter 3
#
Page 25
The buttons provided on Systems Settings Menu #2 are:
PAUSE AFTER
MEASUREMENT:
After each measurement of a manual leakage test, ES601-US will pause for
the indicated time before proceeding with the next measurement of the
sequence. The available settings are "NO" (no pause); 1 second (factory
default); 2, 3, 4 and 5 seconds. The delay provided by this control allows the
user to more easily interrupt a leakage sequence in progress, so a particular
measurement of interest can be repeated as desired.
POWER UP DELAY:
During a manual leakage test, ES601-US will wait for the indicated time when
the AC supply is applied to the DUT, before proceeding with the test
sequence. This delay allows the DUT to start up and stabilize before
subsequent leakage measurements are taken. The available options are 2
seconds (factory default), or 3 to 9 seconds in one second steps. Equipment
requiring more than 9 seconds to start up are better accommodated through
the "STOP AFTER..." control, described below.
STOP AFTER
SUPPLY ON:
If set to "YES", this control will override the power-up delay setting
described above and will cause a manual leakage test to
fully stop
immediately after the AC supply is connected to the DUT by ES601-US.
ES601-US will then prompt the user to activate the DUT and will not
proceed until the user acknowledges by pressing the
CONTINUE
button.
This feature accommodates equipment with a momentary ON/OFF power
switch, which cannot be switched ON until a full AC supply is provided to
the equipment. The factory default setting for this control is "NO", which
then enables the power-up delay described above.
STOP BEFORE
SUPPLY OFF:
If set to "YES", this control will cause a manual leakage test to
fully stop
immediately before a test step which disconnects the AC supply from the
DUT. ES601-US will then prompt the user to shut down the DUT and will
not proceed until the user acknowledges by pressing the
CONTINUE
button.
After
CONTINUE
is pressed, ES601-US will then disconnect the AC supply
from the DUT. This feature accommodates equipment having special
shutdown requirements, such as diagnostic imaging equipment or equipment
incorporating a personal computer. The factory default setting for this control
is "NO".
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...