ES601-US OPERATING MANUAL
Manual Testing/Chapter 5
#
Page 54
Figure 29
- Sequence Controls for manual PE test.
5.9 Ground Resistance Test
Pressing the
GROUND RESISTANCE
button on the MANUAL MODE menu allows you to measure
the resistance of the DUT's ground circuit, using a test current of 1 AMP.
Upon selecting this test from the menu, ES601-US will recall the Systems Settings for the manual ground
resistance test and will then display these on the LCD:
You may change these settings to suit the upcoming test. After the test is completed, the Sequence Controls
will revert to their System Setting defaults.
The buttons provided on the Sequence Control menu are:
TEST MULTIPLE
RESISTANCES:
If set to "YES", this control will enable multiple ground and/or point-to-point
resistance measurements within the scope of the test. Up to ten resistance
measurements may be recorded by ES601-US in multiple resistance mode. If
set to "NO", the test will be completed after performing a single measurement
of resistance.
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...