ES601-US OPERATING MANUAL
Automated Testing/Chapter 6
#
Page 125
Figure 92
- Checklist in progress, M (Tester) Checkitem.
If the command of the M Checkitem is not one of the predefined commands in the table above, ES601-US
will then transmit the command via the RS232 Tester Port specified in the AutoSequence setup. The
prompt portion of the Checkitem will then be displayed on the LCD as the "TASK". However, if the first
character of the Checkitem command is a whitespace or a <TAB> character, ES601-US does not transmit
anything and will immediately display the user interface of
Figure 90
.
In the event that data is not immediately received from the tester in response to the command, a count-up
timer is displayed on the "TEST DATA" line, as shown above. Once test data is received, the timer is
removed from the LCD, replaced by the received data. While waiting for data, or after data is received,
PASS
,
FAIL
,
BACK
and
SKIP
buttons are active and take their usual functions (see P Checkitem).
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...