ES601-US OPERATING MANUAL
Manual Testing/Chapter 5
#
Page 74
Figure 42
- Switch DUT to "ON" for upcoming "neutral open" tests.
After
CONTINUE
is pressed, ES601-US will open the Neutral and will then prompt you to reset the DUT
power switch to "ON". This is necessary to satisfy the requirements of the leakage test, but the instruction
can be ignored if the DUT is of the type that cannot be set to "ON" unless a full AC supply is provided.
Note the purpose of the "STOP BEFORE..." break is to allow you to accommodate power-sensitive
equipment, such as a personal computer or equivalent DUT, which cannot be arbitrarily disconnected from
the AC supply without potentially damaging the DUT. This type of equipment is much less common than
equipment that can be switched "OFF" at any time. The more typical setting of "STOP BEFORE..." is
therefore "NO", in which case the BREAK menus of
Figure 39
and
Figure 40
will not be displayed.
Many types of medical equipment do not fully disconnect from the AC supply when the power switch is set
to "OFF". Even when "OFF", this type of equipment remains effectively connected to the AC supply and
continues to operate in a standby state, for example, by charging an internal battery. For this type of
equipment, "DUT OFF" leakage tests are required under AAMI-ES1 and IEC60601 standards.
"DUT OFF" leakage measurements are optionally enabled by the user before the test is performed, by
means of the leakage setup menu. If "DUT OFF" tests are enabled, ES601-US will stop after all leakage
measurements have been completed with DUT power switch in "ON" position. ES601-US will then display
instructions as shown in the following figure.
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...