ES601-US OPERATING MANUAL
Manual Testing/Chapter 5
#
Page 55
TEST
GROUND WIRE
RESISTANCE:
If set to "YES", this control will enable measurement of the ground wire
(power cord) resistance. For this type of test, resistance is measured between
the ground terminal on the power cord connected to the test receptacle of
ES601-US, and the Kelvin Cable which will be connected to a grounding stud
or the metal enclosure of the DUT.
TEST
POINT-TO-PT
RESISTANCE:
If set to "YES", this control will enable measurement of resistance between two
points, one being the point of attachment of the Kelvin Cable, and the other
being the point of attachment of a test lead which connects to the "TO DUT
CASE" jack of ES601-US. If set to "NO", point-to-point resistance testing is
disabled. This control works with the other two controls to provide various
protocols for ground resistance testing as shown in the table below.
MULTI
GND
WIRE
POINT-
TO-PT
Test Protocol
Test Application
NO
YES
NO
Measure resistance once, between
Kelvin Cable and test receptacle of
ES601-US
.
Conventional ground resistance test of
equipment which connects to the AC
supply by means of a power cord.
NO
NO
YES
Measure resistance once, between
Kelvin Cable and "DUT CASE" lead.
The
OFFSET
calibration is subtracted
from the measurement to obtain the
final result.
The Kelvin Cable is connected to an
equipotential (grounding) stud located
near the DUT, and the "DUT CASE"
lead connects to the DUT itself.
Single ground resistance test of
permanently-installed equipment which
is directly wired to the AC supply.
Example: test of X-RAY equipment.
YES
YES
NO
Measure resistance up to 10 times,
between Kelvin Cable and test
receptical of
ES601-US
.
For the first measurement, the Kelvin
Cable is connected to a designated
ground test stud on the equipment
chassis, per the standard test for
power cord-connected equipment.
For subsequent tests, the Kelvin Cable
may be re-attached at arbitrary
locations on the equipment enclosure
to confirm continuity of the ground to
those points.
Multiple ground resistance test of cord-
connected equipment.
Example: test of patient monitor system.
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...