ES601-US OPERATING MANUAL
Manual Testing/Chapter 5
#
Page 49
Figure 25
- Load current result.
After the load current is displayed, the following options become available:
CONTINUE:
Copies the displayed value to the temporary data buffer. Previous load current data in
the buffer, if any, are overwritten. Returns to the MANUAL MODE menu.
REPEAT:
Repeats the Load Current measurement.
ESCAPE:
Returns to the MANUAL MODE menu. The result is discarded and is not copied to
the data buffer.
After pressing either
REPEAT
or
CONTINUE
, and if "STOP BEFORE..." of the System Settings is
"NO" (see Section 3.3), ES601-US will switch OFF the test socket and return to the MANUAL MODE
menu. If "STOP BEFORE..." is "YES", ES601-US will prompt you to shut down the equipment before the
test socket is switched OFF:
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...