ES601-US OPERATING MANUAL
ES601pc-US Software/Chapter 8
#
Page 165
8.8.5 Deleting a Sequence
An entire sequence can be deleted by pressing the "Delete Sequence" button. A message box will appear
prompting you to be sure you want to delete the sequence. Press "Yes" if you are sure, and all the
waveforms in the sequence will be removed as well as the sequence title.
8.8.6 Uploading AutoSequences
To upload ECG AutoSequences to ES601-US, press the "Upload To ES601" button. Choose either
uploading via USB, Ethernet, or Serial Port, and then press "Connect". If the connection is via ethernet,
enter the IP address assigned to ES601-US (the IP address of ES601-US is displayed on the LCD when the
analyzer is operating in Ethernet Mode). If the connection is via serial port, choose the COM port to which
the ES601-US is connected (see
Figure 108
, under Safety AutoSequences).
After pressing "Connect", a message box with a transfer progress bar will appear (see
Figure 109
, under
Safety AutoSequences). After the transfer is completed, a Transfer Complete message will appear and the
message box will disappear.
8.8.7 Downloading AutoSequences
Sometimes it is desirable to bring data stored in the ES601-US back into your PC. This way, you can
transfer AutoSequence setups from one ES601-US analyzer to another.
You can retrieve AutoSequence setups from a ES601-US by pressing the "Download From ES601" button.
Choose to download via USB, Ethernet, or Serial Port. Doing so will overwrite the configuration of all 24
Safety AutoSequences in ES601pc-US with the sequences from the ES601-US.
These sequences can then be uploaded to a different ES601-US, or they can be optionally modified and
uploaded to the same ES601-US unit, or additional ES601-US units.
8.8.8 Printing AutoSequences
Press the Print button to show the print dialogue. Select the current sequence radio button and press OK to
print out the current sequence. Select the All Sequences radio button and press "OK" to print out all the
sequences. Pressing "Cancel" will close the print dialogue without printing.
8.8.9 Return to Main Window
When edits have been completed, the user may return to the main menu of ES601pc-US by pressing this
button. When this button is pressed, all changes which have been made will be saved on your computer.
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...