ES601-US OPERATING MANUAL
Manual Testing/Chapter 5
#
Page 70
Figure 38
- Leakage current test in progress.
WARNING:
If testing of
Type BF
or
Type CF
patient leads has been enabled, HIGH VOLTAGE (isolation test
voltage) will be periodically output from the AP jacks during the test.
The isolation test voltage is isolated from the AC mains and is internally current-limited to 8 milliamperes
maximum at 264 VAC output. Unterminated leads attached to the AP jacks therefore present a minor,
non-lethal shock hazard to the careless user.
Always take care to make proper electrical connections to the DUT before starting a leakage test.
Disconnect all unused leads or cables from the AP jacks on ES601-US.
Once the leakage test has been set up and the connections are made, the test is started by pressing the
CONTINUE
button on the CONNECTIONS menu. The required measurements are performed
automatically in a sequence. AC RMS and DC currents are measured simultaneously, and the display is
updated with new data approximately two times every second. If the "PAUSE AFTER..." Sequence Control
is not set to "NO", ES601-US will delay an additional 1 to 5 seconds between consecutive leakage
measurements.
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...