ES601-US OPERATING MANUAL
Appendix A
#
Page 199
Insulation Resistance - Hot+Neutral to Ground: Skip
Insulation Resistance - Patient Leads to Ground: Skip
Ground Resistance: Measured At 1 Amp
Leakage Current: Measure
Equipment Power: On/Off
Polarity: Normal/Reverse
Neutral: Normal/Open
Ground: Normal/Open
Test External and Internal Case Leakages, plus:
Patient Leads
Lead to Lead Leakages
Patient Lead Groups
Group 1:
Name: ECG
Type: Isolated
Leads: 4
LD 1 - RA
LD 2 - RL
LD 3 - LA
LD 4 - LL
SAFETY AUTOSEQUENCE 4 (SAF04)
A4,A9: CASE X,I
CASE LKG (CRITICAL)
Standard: AAMI-ES1
Limits Used: AAMI-ES1 Limits
Equipment Class: Class I
Line Voltage: Measure
Load Current: Measure
Insulation Resistance - Hot+Neutral to Ground: Skip
Insulation Resistance - Patient Leads to Ground: Skip
Ground Resistance: Measured At 1 Amp
Leakage Current: Measure
Equipment Power: On/Off
Polarity: Normal/Reverse
Neutral: Normal/Open
Ground: Normal/Open
Test External and Internal Case Leakages
Patient Lead Groups
None
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...