4.2 Connecting a Battery-Powered DUT to ES601-US
Connection instructions for battery-powered DUT:
• Whenever connecting or disconnecting a DUT or peripheral devices to ES601-US, the power
switches of all equipment should be OFF.
• The “TO DUT CASE” jack of ES601-US connects to the enclosure of the DUT, or alternatively, to
aluminum foil which is wrapped over the DUT if the enclosure is made of plastic.
• As required, AUX1 may connect to electrically isolated input/output signals. Most typically, this
connection would be for testing a transformer-isolated data communications interface on the DUT,
such as an Ethernet Port.
• Note the Test Receptacle and the Kelvin Cable are not used for testing internally-powered devices.
Summary of Contents for ES601-US
Page 1: ...Operating Manual ES601 US Automated Electrical Safety Analyzer ...
Page 2: ......
Page 4: ......
Page 8: ...Page iv ...
Page 20: ...ES601 US OPERATING MANUAL Definitions Page xvi ...
Page 40: ...ES601 US OPERATING MANUAL Overview Chapter 2 Page 20 ...
Page 58: ...ES601 US OPERATING MANUAL Setup Chapter 3 Page 38 ...
Page 114: ...ES601 US OPERATING MANUAL Manual Testing Chapter 5 Page 94 ...
Page 135: ...ES601 US OPERATING MANUAL Automated Testing Chapter 6 Page 115 ...
Page 162: ...ES601 US OPERATING MANUAL Computer Control Chapter 7 Page 142 ...
Page 216: ...ES601 US OPERATING MANUAL Maintenance Chapter 9 Page 196 ...
Page 234: ...ES601 US OPERATING MANUAL Appendix B Page 214 ...
Page 238: ...ES601 US OPERATING MANUAL Appendix C Page 218 ...
Page 239: ...ES601 US OPERATING MANUAL Appendix D Page 219 APPENDIX D Importing Equipment Records ...
Page 240: ...ES601 US OPERATING MANUAL Appendix D Page 220 ...