Q00110 Rev A 8/06
Page 88 of 225
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
7.
Using the alpha-numeric keyboard, enter in the required TCP/IP network configuration
values/settings in the appropriate fields, as specified in the information previously provided in
the “Pre-Installation Site Survey Form”, by the medical facility’s IT System Administrator
.
8.
Using the
Menu Control
, arrow down to backlight
Apply
selection in the main menu, and
press the SELECT button
to save the serial port configuration setup, and exit.
9.3 “Serial Port” Setup
The “SERIAL PORT” configuration menu allow for configuring the operating parameters (BAUD
rate, start/stop bits, etc) applied in use of (one) the serial ports at the Rear I/O panel of the SuperCart.
The serial port(s) will be used (
in the future
) by the Operator for the purpose of outputting raw OB
calculation report table data from the Z.ONE, to remote report generator devices.
1.
Using the page TAB button
to advance menu screen to select
TOOLS
menu.
TOOLS
Menu
2.
Using the
Menu Control
, arrow down to backlight
SERIAL PORT
selection, and press the
SELECT button
to bring up this menu.