Unable to modify existing User Preset (shown as grayed out on screen)
•
Ensure that the desired user preset has been previously (prior to attempting to modify it)
selected as the active preset.
19.3.8 DICOM Connectivity/Communications Problems
Z.ONE
will not communicate with target DICOM” Store” or “Print” device
•
Check Networking ICON
on lower/right corner of main imaging screen of
Z.ONE
, to
verify hardware connection is active (no “
X
” over ICON). Check hardware cabling, network
equipment, and/or configuration, as necessary.
•
In
General
” configuration page, verify DICOM Application Title, and IP
address/gateway/netmask parameters are entered correctly, and correspond with those assigned
to
z.one
ultrasound system by the IT Admin at the install site.
•
In the DICOM”
Network Store
” and “
Printer
” configuration pages, verify AE Title, IP
Address, and Port Number parameters are entered correctly, and match those assigned to the
corresponding device by the IT Admin for facility, at install site.
•
Using the “VERIFY” selection in the DICOM configuration menus, for the Network Storage
or Printer Administration screens (as applicable to device attempting to be connected), test for
successful handshaking of connection.
•
Using another device (PC, etc.) on same domain on hospital network, perform a “PING”
operation to the IP address assigned for the
z.one
, in the “General” DICOM configuration
page, to test for TCP/IP connectivity.