NOTE 1: If the cable has
too much
slack (the lift is not properly releasing), rotate the
adjuster barrel in the
counter-clockwise
(as viewed from the top) direction, a
sufficient amount to reduce cable slack, and provide an optimum release point on
the release lever handle.
NOTE 2: If the cable is
too tight
(the lift is not staying locked in position when the release
handle is totally released – aka “drifting”), rotate the adjuster barrel in the
clockwise
(opposite to what is shown in the illustration above) direction just far
enough to resolve the “drifting” condition.
3.
Reverse cable access process above, to complete this procedure.
18.5
Touchscreen Calibration - (Scan Engine)
IMPORTANT
•
The touchscreen is very sensitive to the “technique” used by the Operator in making
contact at each of the calibration points with the stylus. It is very important that only a
single discrete (and firm) contact be made at each of the green splash graphics
calibration points. Any excessive debounce, when pressing the stylus to the screen, will
likely result in a failure of the calibration process.
18.5.1 Calibration Procedure
1.
Remove the Scan Engine from the SuperCart
2.
Press the On/Off switch on the Scan Engine, to power-on the
z.one
3.
Using the “Tab” button and Menu Control on the Scan Engine user interface, advance to the
“Tools” menu
and then sequence through the following selection:
Tools
tab --> SYSTEM SETUP --> Touchscreen
4.
Press the “SELECT”
button on the user interface. The Touchscreen calibration screen
should appear, with the first (of 3) green “splash” target points being should now be
displayed