Figure 37: AC Power Adapter
STEPS:
1.
Power off the Scan Engine, and ensure it is being used remotely (undocked) from the
SuperCart or
mini
Cart
2.
Plug the small connector of the DC Adapter Cable, into the mating socket on the DC power
port, located at the back of the Scan Engine (inside the Rear I/O access door).
3.
Connect the output connector of the DC Power Cable (out of the Power Supply) to the DC
Adapter Cable’s mating connector
4.
Plug the AC Power Cable’s female connector into the socket on the Power Supply
5.
Plug the other end of the AC Power Cable into an active AC wall outlet
6.
Inspect the Power-On indicator LED, on the Power Supply brick, to ensure power is on
7.
Power on the Scan Engine and begin normal operation
5.6.8
2-Bay Battery Charger: Use & Operation
WARNING
There are no User or Field Service repairable components within the 2-Bay Battery Charger.
Any defective units should be returned to the Zonare factory for repair/replacement.
The Lithium-Ion battery packs, used to power the Scan Engine during standalone (undocked) use,
have expected charge retention of approximately 1 hour (during full-use of the Scan Engine). The
battery pack is normally kept charged when the Scan Engine is docked within the SuperCart.
However to prepare multiple battery packs for standalone Scan Engine use, or to recalibrate battery
packs for maximum performance, a Zonare 2-Bay Battery Charger is available.