Q00110 Rev A 8/06
Page 179 of 225
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
Nut-Driver Set (metric)
Multi-sizes
1 set
5.5mm needed for AC Outlet box
Combination Wrench Set, small sizes (metric)
Multi-sizes
1 set
General use
Flashlight, Mini-Mag
6” size
1
Illuminating tight access areas
Magnetic pick-up tool
Telescoping
1
Retrieval of hardware
20.3 Hardware Service/Replacement Procedures (SuperCart)
20.3.1 Main - I/O Board Assembly - Removal/Replacement (SuperCart)
Required Parts
•
P/N: 80207............... Assy, SuperCart Main Board w/IO & Panel
Required Tools
•
ESD protection equipment (wrist grounding strap, anti-static surface)
•
Philips head screwdriver
IMPORTANT NOTE
Due to requirements for matching the video output signal of the Main-I/O board to the
calibration settings in the 13” LCD (Barco) Display monitor, it is required to perform the “Full
Autoset” portion of the 13” LCD (Barco) Display calibration procedure (provided in the
previous chapter of this manual), anytime the Cart Main-I/O board is replaced in the field.
Procedure
STEPS:
1.
Ensure the
z.one
is powered “OFF”
2.
Place AC circuit breaker, located at the rear of the SuperCart, in the “OFF” (down) position
3.
Disconnect the main AC power cord from rear of the SuperCart
4.
Disconnect all external cables (network, video, etc.) from the I/O panel at the rear of the
SuperCart
5.
Using a Philips head screwdriver, loosen the two (2) thumbscrews that secure the Main-I/O board
assembly to the chassis