Q00110 Rev A 8/06
Page 28 of 225
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
•
Power for peripherals: 300W
3.10 System Power Protection
•
SuperCart
-
Re-settable AC circuit breaker
•
miniCart
-
User-replaceable fuses:
−
T6.3A/250V ..................................................... (100-120V Systems)
−
T3.15A/250V ................................................... (200-240V Systems)
3.11 Electrical Specifications
DC Voltages:
•
SuperCart DC Power Supplies
Voltage
Purpose
Max Current Rating
+12V
SuperCart Main Logic / Scan Engine Power
20 A
+12V
LCD Display Source Power
6 A
+5V
Sup5V Logic Power
8 A
•
miniCart DC Power Supplies
Voltage
Purpose
Max Current Rating
+12V
Scan Engine Power
20 A
+12V
LCD Display Source Power
6 A
+5V
USB Hub Power
8 A