Q00110 Rev A 8/06
Page 203 of 225
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
3.
The fuse holder is located directly above the connector (at the lower-left, rear of the miniCart)
where the main AC power cord plugs into the back of the miniCart.
4.
Using a small screwdriver (or the edge of a fingernail, if applicable), lift up on the plastic release
tab (shown by the arrow below), in the slot at the bottom-center of the fuse holder’s cover plate.
5.
Slide the fuse holder assembly out of its receptacle.
6.
Remove both fuses from the fuse holder, and inspect.
7.
Replace fuses as required, and reverse above steps for re-installation.
Fuse Specifications:
- 100-120V Systems: T6.3A/250V
- 200-240V Systems: T3.15A/250V
20.4.2 Main Power Supply – Removal/Replacement (miniCart)
Required Parts
•
P/N: 32621-00 .......
Power Supply , 250W, 5V@8A, 12V@6A, 12V@20A, Med. Grade
Required Tools/Equipment
•
#2 Philips head screwdriver