Q00110 Rev A 8/06
Page 74 of 225
"Note: Copies are uncontrolled documents - For revision verification see the Master Documentation List"
8.2 “DICOM” Configuration Parameter Definitions: (
Z-ONE
)
The table below provides some guidance regarding the type of information required for the DICOM
configuration of the various menus (“
General
”, “
Printers
”,
Network Store
”, “
MPPS
” and
“
Worklist
”) on the
z.one
.
DICOM Configuration Parameters:
Definition:
AE Title
e.g., “Zonare1”
The official “name” (
Application Entity
) assigned to the Z.ONE (by the network
administrator) to identify it to the network.
Compression
−
None
−
RLE Lossless
−
JPEG Lossy
Enables the ability to compress image files for the purpose of reducing file size
and transfer times.
RLE Lossless automatically reduces the size of the transferred image, yet retains
all source content.
Provide the highest level of image file size reduction (compression) of all the
options. User selectable “Quality” setting (between 85-100%) determines
resultant file size.
Date Range
Specify the number of days back (Today, 2, 7, All) that the DICOM worklist will
reference scheduled examinations.
Gateway IP Address
A designated TCP/IP address (i.e 10.1.1.254) that is required in cases where the
Z.ONE
is to be connected to target devices that are outside the primary host
network (FTP remote access, for example).
ICON Image
A sub-sampled (miniature size) version of the image created for each patient
image on the
Z.ONE
. Options are to either
retain
or
remove
these images from
DICOM image transfers to the target device. (default is “
retain
”)
IP address
Internet Protocol address. A unique Internet address for devices attached to an IP
network (TCP/IP network).
Keep Alive Timeout
Specify the duration (in seconds) that an established DICOM connection will be
maintained, after completion of the last transfer..
Limit by Modality
−
None
−
US (ultrasound)
−
50+ additional options:
(see list of selections)
Filter the Worklist by modality (US, CT, MRI, et cetera). Set according to
department requirements.
None: Show entries for
all
modalities.
US: Show entries for
ultrasound
exams only.
The following additional modality filter (limit) settings are available for selection:
Selections – AU, BI, CD, CR, CT, DD, DG, DX, ECG, EPS, ES, FID, GM, HC, HD,
IO, IVUA, KO, LS, MG, MR, NM, OP, OT, PR, PT, PX, REG, RF, RG, RTDOSE,
RTIMAGE, RTPLAN, RTRECORD, RWV, SM, SMR, SR, ST, TG, US, XA, XC
Log Level
Defines the level of detail information that will be displayed in the DICOM
Queue.
Maximum entries cached
Enter a number between 0 and 999
Set according to department requirements. A higher number provides access to
more scheduled exams, but increases the time required to search for the exam of
interest.
Minimum/Maximum Film Density
For defining Minimum/Maximum density of images on film. Value entered is
expressed in hundredths of OD (optical density). A value entered of 150,
corresponds with 1.5 OD. Typical acceptable values to be entered are 000-399,
Содержание Z.ONE