INSTrument:CREate[:NEW]
<ChannelType>, <ChannelName>
This command adds an additional measurement channel. You can configure up to 10
measurement channels at the same time (depending on available memory).
See also
●
●
Parameters:
<ChannelType>
Channel type of the new channel.
For a list of available channel types see
<ChannelName>
String containing the name of the channel.
Note that you can not assign an existing channel name to a new
channel; this will cause an error.
Example:
INST:CRE SAN, 'Spectrum 2'
Adds an additional spectrum display named "Spectrum 2".
Manual operation:
See
INSTrument:CREate:REPLace
<ChannelName1>,<ChannelType>,<ChannelName2>
This command replaces a channel setup with another one.
Setting parameters:
<ChannelName1>
String containing the name of the channel setup you want to
replace.
<ChannelType>
Channel type of the new channel setup.
For a list of available channel setup types see
<ChannelName2>
String containing the name of the new channel setup.
Note
: If the specified name for a new channel setup already
exists, the default name, extended by a sequential number, is
used for the new channel setup (see
on page 608).
Channel names can have a maximum of 31 characters, and
must be compatible with the Windows conventions for file
names. In particular, they must not contain special characters
such as ":", "*", "?".
Example:
INST:CRE:REPL 'IQAnalyzer2',IQ,'IQAnalyzer'
Replaces the channel setup named "IQAnalyzer2" by a new
channel setup of type "IQ Analyzer" named "IQAnalyzer".
Usage:
Setting only
Manual operation:
See
"Replace Current Channel Setup"
Selecting the Operating Mode and Application