General Instrument Setup
R&S
®
FPL1000
545
User Manual 1178.3370.02 ─ 11
Remote command:
Abort Self Alignment
As long as the self-alignment data is being collected, the procedure can be canceled
using the "Abort Self Alignment" button.
Note:
If you start a self-alignment remotely, then select the "Local" softkey while the
alignment is still running, the instrument only returns to the manual operation state
after the alignment is completed. In this case, you cannot abort a self-alignment man-
ually.
Await Warm-Up Operation before Self Alignment
Displays a message indicating the remaining warmup time required before self-align-
ment is performed. After the warmup operation has completed, self-alignment is star-
ted automatically.
Schedule
If enabled, a self-alignment is performed regularly at specific days and time.
Enter the time according to the format specified in
Shut down Device after Self Alignment
If activated, the R&S
FPL1000 is automatically shut down after self-alignment is com-
pleted. Note that the instrument cannot be restarted via remote control.
Starting Touch Screen Alignment
Starts the touchscreen alignment.
Tap the 4 markers on the screen as you are asked to do. The touchscreen is aligned
according to the executed pointing operations.
Alignment Results:
Information on whether the alignment was performed successfully and on the applied
correction data is displayed. The results are available until the next self-alignment
process is started or the instrument is switched off.
Alignment