Remote Commands
R&S
®
FPL1000
768
User Manual 1178.3370.02 ─ 11
Parameters:
<Attenuation>
Range:
see data sheet
Increment: 5 dB (with optional electr. attenuator: 1 dB)
*RST:
10 dB (AUTO is set to ON)
Default unit: DB
Example:
INP:ATT 30dB
Defines a 30 dB attenuation and decouples the attenuation from
the reference level.
Manual operation:
See
INPut<ip>:EATT:AUTO
<State>
INPut<ip>:ATTenuation:AUTO
<State>
This command couples or decouples the attenuation to the reference level. Thus, when
the reference level is changed, the R&S
FPL1000 determines the signal level for opti-
mal internal data processing and sets the required attenuation accordingly.
For the R&S
FPL1000, these commands are identical.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
INP:ATT:AUTO ON
Couples the attenuation to the reference level.
Manual operation:
See
10.8.3.3
Configuring a Preamplifier
INPut<ip>:GAIN:STATe
<State>
This command turns the internal preamplifier on and off. It requires the optional pream-
plifier hardware.
If activated, the input signal is amplified by 20
dB. The preamplifier is only active below
3
GHz (R&S
FPL1003) or 7.5
GHz (R&S
FPL1007).
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off