Remote Commands
R&S
®
FPL1000
793
User Manual 1178.3370.02 ─ 11
Example:
SOUR:INT:FREQ:OFFS 10kHz
Manual operation:
See
[SENSe:]CORRection[:STATe]
<State>
This command turns correction of measurement results (normalization) on and off.
The command is available after you have created a reference trace for the selected
measurement type with
[SENSe:]CORRection:COLLect[:ACQuire]
This command is only available if the Tracking Generator is active (see
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
1
Example:
CORR ON
Activates normalization.
Manual operation:
See
[SENSe:]CORRection:COLLect[:ACQuire]
<MeasType>
This command initiates a reference measurement (calibration). The reference mea-
surement is the basis for the measurement normalization. The result depends on
whether a reflection measurement or transmission measurement is performed (see
To obtain a correct reference measurement, a complete sweep with synchronization to
the end of the sweep must have been carried out. This is only possible in the single
sweep mode.
This command is only available if the Tracking Generator is active (see
Setting parameters:
<MeasType>
THRough | OPEN
THRough
"TRANsmission" mode: calibration with direct connection
between generator and device input
"REFLection" mode: calibration with short circuit at the input
OPEN
only allowed in "REFLection" mode: calibration with open input
Setting Basic Spectrum RF Measurement Parameters