Remote Commands
R&S
®
FPL1000
761
User Manual 1178.3370.02 ─ 11
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
BAND:VID:AUTO OFF
Manual operation:
See
See
See
[SENSe:]BWIDth:VIDeo:RATio
<Ratio>
[SENSe:]BANDwidth:VIDeo:RATio
<Ratio>
This command defines the coupling ratio of the video bandwidth to the resolution band-
width (RBW/VBW).
Parameters:
<Ratio>
Range:
0,001 to 1000
*RST:
1
Example:
BAND:VID:RAT 3
Sets the coupling of video bandwidth to video bandwidth =
3*resolution bandwidth
Manual operation:
See
[SENSe:]BWIDth:VIDeo:TYPE
<Mode>
[SENSe:]BANDwidth:VIDeo:TYPE
<Mode>
This command enables or disables the logarithmic amplifier in front of the video filter in
the signal path.
Parameters:
<Mode>
LINear
The logarithmic amplifier in front of the video filter is bypassed to
process linear detector samples.
LOGarithmic
The logarithmic amplifier in front of the video filter is enabled to
process logarithmic detector samples.
*RST:
LOGarithmic
Example:
BAND:VID:TYPE LIN
Logarithmic amplifier in front of the video filter is disabled.
10.8.2.2
Configuring the Sweep
Useful commands for configuring sweeps described elsewhere:
●
●
●
Setting Basic Spectrum RF Measurement Parameters