![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 821](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987821.webp)
Remote Commands
R&S
®
FPL1000
821
User Manual 1178.3370.02 ─ 11
LINear
Linear calculation.
LOGarithmic
Logarithmic calculation.
POWer
Linear power calculation.
*RST:
LOGarithmic
Example:
CALC:MATH:MODE LIN
Selects linear calculation.
Manual operation:
See
CALCulate<n>:MATH<t>:POSition
<Position>
This command defines the position of the trace resulting from the mathematical opera-
tion.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Position>
Vertical position of the trace in % of the height of the diagram
area.
100 PCT corresponds to the upper diagram border.
Range:
-100 to 200
*RST:
50
Default unit: PCT
Example:
CALC:MATH:POS 100
Moves the trace to the top of the diagram area.
Manual operation:
See
CALCulate<n>:MATH<t>:STATe
<State>
This command turns the trace mathematics on and off.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Setting Basic Spectrum RF Measurement Parameters