![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 854](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987854.webp)
Remote Commands
R&S
®
FPL1000
854
User Manual 1178.3370.02 ─ 11
<m>
Manual operation:
See
"Search Mode for Next Peak in Y-Direction"
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:NEXT
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum:NEXT
This command moves a marker vertically to the next lower peak level for the current
frequency.
The search includes all frames. It does not change the horizontal position of the
marker.
Suffix:
<n>
.
<m>
Manual operation:
See
"Search Mode for Next Peak in Y-Direction"
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum[:PEAK]
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum[:PEAK]
This command moves a marker vertically to the highest level for the current frequency.
The search includes all frames. It does not change the horizontal position of the
marker.
If the marker hasn't been active yet, the command looks for the peak level in the whole
spectrogram.
Suffix:
<n>
.
<m>
CALCulate<n>:MARKer<m>:SGRam:Y:MINimum:ABOVe
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MINimum:ABOVe
This command moves a marker vertically to the next higher minimum level for the cur-
rent frequency.
The search includes only frames above the current marker position. It does not change
the horizontal position of the marker.
Suffix:
<n>
.
<m>
Manual operation:
See
"Search Mode for Next Peak in Y-Direction"
Setting Basic Spectrum RF Measurement Parameters