![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 20](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987020.webp)
Getting Started
R&S
®
FPL1000
20
User Manual 1178.3370.02 ─ 11
●
The R&S
FPL1000 allows for battery operation if option R&S
FPL1-B31 is installed.
4.1.5.1
Connecting the AC Power
The R&S
FPL1000 can be used with different AC power voltages and adapts itself
automatically to it. Refer to the datasheet for the requirements of voltage and fre-
quency.
For safety information, see
To connect the AC power
1. Plug the AC power cable into the AC power connector on the rear panel of the
instrument. Only use the AC power cable delivered with the R&S
FPL1000.
2. Plug the AC power cable into a power outlet with ground contact.
The required ratings are listed next to the AC power connector and in the data
sheet.
For details on the connector, refer to
Chapter 4.2.2.1, "AC Power Supply Connection
4.1.5.2
Connecting an Optional DC Power Supply (R&S
FPL1-B30)
The R&S
FPL1000 can also be equipped with an optional DC power supply connector
(R&S
FPL1-B30). If installed, the R&S
FPL1000 can be operated by a DC voltage of
+12
V to +24
V. For details on the connector see
Chapter 4.2.2.2, "Li-Ion Battery Packs
If you use an external power supply unit to supply safety extra-low DC voltage (SELV)
to the instrument, be sure to meet the requirements for reinforced/double insulation in
accordance with DIN/EN/IEC 61010 (UL 3111, CSA C22.2 No. 1010.1) or DIN/EN/IEC
60950 (UL 1950, CSA C22.2 No. 950). Provide current limitation in accordance with
DIN EN 61010-1 Appendix F2.1. Use a cable no longer than 3
m.
Also see
Preparing for Use