Getting Started
R&S
®
FPL1000
38
User Manual 1178.3370.02 ─ 11
The meanings of the labels on the R&S
4.2.2.1
AC Power Supply Connection and Main Power Switch
An AC power supply connector and main power switch are located in a unit on the rear
panel of the instrument.
Main power switch function:
Position 1: The instrument is in operation.
Position O: The entire instrument is disconnected from the AC power supply.
4.2.2.2
Li-Ion Battery Packs and DC Power Connector
With the Li-ion battery pack (option R&S
FPL1-B31), the R&S
FPL1000 can be oper-
ated independently of an AC or DC power supply. The instrument can house 2 Li-ion
batteries which can be charged both via AC or DC power supply.
For safety information concerning batteries, see
As an alternative, a DC power supply connector (option R&S
FPL1-B30) is available.
DC power supplies from +12
V to +24
V and from 13
A to 6.5
A can be used. Connect
the connector according to the following diagram:
Pin
Description
1
2
3
1
Plus
2
Ground
3
Not used
If you use an external power supply unit to supply safety extra-low DC voltage (SELV)
to the instrument, be sure to meet the requirements for reinforced/double insulation in
accordance with DIN/EN/IEC 61010 (UL 3111, CSA C22.2 No. 1010.1) or DIN/EN/IEC
60950 (UL 1950, CSA C22.2 No. 950). Provide current limitation in accordance with
DIN EN 61010-1 Appendix F2.1.
Also see
4.2.2.3
GPIB Interface
The optional GPIB interface (R&S
FPL1-B10) is in compliance with IEEE488 and
SCPI. A computer for remote control can be connected via this interface. To set up the
connection, a shielded cable is recommended.
Instrument Tour