![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 873](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987873.webp)
Remote Commands
R&S
®
FPL1000
873
User Manual 1178.3370.02 ─ 11
Suffix:
<n>
.
<m>
Parameters:
<Mode>
POWer
Result is displayed as an absolute power. The power unit
depends on the
DENSity
Result is displayed as a density in dBm/Hz.
RPOWer
This setting is only available for a delta band power marker.
The result is the difference between the absolute power in the
band around the delta marker and the absolute power for the
reference marker. The powers are subtracted logarithmically, so
the result is a dB value.
[Relative band power (Delta2) in dB] = [absolute band power
(Delta2) in dBm] - [absolute (band) power of reference marker in
dBm]
*RST:
POWer
Manual operation:
See
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:RESult?
This command queries the results of the band power measurement.
Suffix:
<n>
.
<m>
Return values:
<Power>
Signal power over the delta marker bandwidth.
Usage:
Query only
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:SPAN
<Span>
This command defines the bandwidth around the delta marker position.
Suffix:
<n>
.
<m>
Parameters:
<Span>
Frequency. The maximum span depends on the marker position
and R&S
FPL1000 model.
*RST:
5% of current span
Default unit: Hz
Setting Basic Spectrum RF Measurement Parameters