Data Management
R&S
®
FPL1000
500
User Manual 1178.3370.02 ─ 11
Figure 7-5: Example for measurement-specific information in a test report
State
Enables or disables the user-defined measurement information on the test report.
Remote command:
HCOPy:TREPort:ITEM:HEADer:STATe
Title
Defines a title for the type of information, for example "Test equipment" for the name of
the device under test. Default titles are provided, but you can change and customize
each title, except for the initial "heading". A maximum of 17 characters are available.
For data that is specific to an individual measurement, this title is used in the dialog
box asking you to provide the information when you append the measurement results
to the report.
Remote command:
HCOPy:TREPort:ITEM:HEADer:LINE<line>:TITLe
Value
Defines the actual text for the content defined by the title. A maximum of 25 characters
are available.
For data that is specific to an individual measurement, you are asked to provide this
information when you append the measurement results to the report.
Remote command:
HCOPy:TREPort:ITEM:HEADer:LINE<line>:TEXT
Visibility
Determines the validity of the content and thus when the contents are configured and
where the line is displayed in the report.
"Never"
(Default): Contents are ignored.
"Global"
Contents are valid globally for the entire report. They are configured
in the template and included on each page.
Working with Test Reports