User Manual 1178.3370.02 ─ 11
8. In the Time Domain Power configuration dialog box, set all four results to "On".
9. Set the "Limit State" to "On".
10. Define the left limit at
326
μ
s
and the right limit at
538
μ
s
.
This range corresponds to the useful part of the GSM burst.
The mean power of the useful part of the GSM burst is calculated to be -13 dBm.
6.2.11
Harmonic Distortion Measurement
The "Harmonic Distortion" measurement measures harmonics and their distortion,
including the total harmonic distortion.
........................................................................................231
...................................................................................232
..................................................................................234
Harmonic Distortion Configuration
........................................................................ 235
How to Determine the Harmonic Distortion
...........................................................237
6.2.11.1
About the Measurement
With this measurement it is possible to measure the harmonics easily, for example
from a VCO. In addition, the total harmonic distortion (THD) is calculated.
For measurements in the frequency domain, the Harmonic Distortion measurement
starts with an automatic search for the first harmonic (= peak) within the set frequency
range. The center frequency is set to this frequency and the reference level is adjusted
accordingly.
For measurements in zero span, the center frequency remains unchanged.
Measurements and Results