The Spectrum Application (RF Measurements)
R&S
®
FPL1000
416
User Manual 1178.3370.02 ─ 11
Note:
Sequencer. If the Sequencer is active, the "Continuous Sweep" softkey only con-
trols the sweep mode for the currently selected channel setup. However, the sweep
mode only takes effect the next time the Sequencer activates that channel setup, and
only for a channel-defined sequence. In this case, a channel setup in continuous
sweep mode is swept repeatedly.
Furthermore, the [RUN CONT] key controls the Sequencer, not individual sweeps.
[RUN CONT] starts the Sequencer in continuous mode.
For details on the Sequencer, see
Chapter 5.4.1, "The Sequencer Concept"
Remote command:
Single Sweep / Run Single
After triggering, starts the number of sweeps set in "Sweep Count". The measurement
stops after the defined number of sweeps has been performed.
While the measurement is running, the "Single Sweep" softkey and the [RUN SINGLE]
key are highlighted. The running measurement can be aborted by selecting the high-
lighted softkey or key again.
Note:
Sequencer. If the Sequencer is active, the "Single Sweep" softkey only controls
the sweep mode for the currently selected channel setup. However, the sweep mode
only takes effect the next time the Sequencer activates that channel setup, and only for
a channel-defined sequence. In this case, the Sequencer sweeps a channel setup in
single sweep mode only once.
Furthermore, the [RUN SINGLE] key controls the Sequencer, not individual sweeps.
[RUN SINGLE] starts the Sequencer in single mode.
If the Sequencer is off, only the evaluation for the currently displayed channel setup is
updated.
For details on the Sequencer, see
Chapter 5.4.1, "The Sequencer Concept"
Remote command:
CALCulate<n>:SPECtrogram:CONTinuous
Clear Spectrogram
Resets the spectrogram result display and clears the history buffer.
This function is only available if a spectrogram is selected.
Remote command:
CALCulate<n>:SPECtrogram:CLEar[:IMMediate]
Color Map Settings
Access
: "Overview" > "Analysis" > "Traces" > "Spectrogram" > "Color Mapping"
or
: [TRACE] > "Spectrogram Config" > "Color Mapping"
For more information on color maps see
on page 410. For details on
changing color mapping settings see
"How to Configure the Color Mapping"