![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 642](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987642.webp)
Remote Commands
R&S
®
FPL1000
642
User Manual 1178.3370.02 ─ 11
ON | 1
Channel power density in dBm/Hz
OFF | 0
Channel power in dBm
*RST:
0
Example:
CALC:MARK:FUNC:POW:RES:PHZ ON
Output of results referred to the channel bandwidth.
Manual operation:
See
"Channel power level and density (Power Unit)"
[SENSe:]POWer:ACHannel:MODE
<Mode>
This command selects the way the R&S
FPL1000 displays the power of adjacent chan-
nels.
You need at least one adjacent channel for the command to work.
Parameters:
<Mode>
ABSolute | RELative
ABSolute
Shows the absolute power of all channels
RELative
Shows the power of adjacent and alternate channels in relation
to the transmission channel
*RST:
RELative
Manual operation:
See
"Absolute and Relative Values (ACLR Mode)"
10.6.4.8
Programming Examples for Channel Power Measurements
The following programming examples are meant to demonstrate the most important
commands to perform channel power measurements in a remote environment.
Example: Configuring and Performing an ACLR Measurement
............................642
Example: Configuring and Performing an ACLR Measurement
In this example we will configure and perform an adjacent-channel power measure-
ment. Note that this example is primarily meant to demonstrate the remote control
commands, it does not necessarily reflect a useful measurement task. For most com-
mon measurement standards, the R&S
FPL1000 performs the measurement optimally
with the predefined settings, without further configuration.
//-----------Preparing the measurement ------------
//Reset the instrument
*RST
//-------------Preparing the measurement---------------------
//Activate adjacent-channel power measurement.
Configuring and Performing Measurements