![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 783](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987783.webp)
Remote Commands
R&S
®
FPL1000
783
User Manual 1178.3370.02 ─ 11
Example:
SYST:COMM:RDEV:PMET2:DEF '','NRP-Z81','',
'123456'
Assigns the power sensor with the serial number '123456' to the
configuration "Power Sensor 2".
SYST:COMM:RDEV:PMET2:DEF?
Queries the sensor assigned to "Power Sensor 2".
Result:
'','NRP-Z81','USB','123456'
The NRP-Z81 power sensor with the serial number '123456' is
assigned to the "Power Sensor 2".
Manual operation:
See
Configuring Power Sensor Measurements
CALibration:PMETer<p>:ZERO:AUTO ONCE
................................................................... 783
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]
............................................................784
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE
....................................... 784
CALCulate<n>:PMETer<p>:RELative:STATe
.....................................................................784
[SENSe:]PMETer<p>:DCYCle[:STATe]
............................................................................. 785
[SENSe:]PMETer<p>:DCYCle:VALue
...............................................................................785
................................................................................... 786
[SENSe:]PMETer<p>:FREQuency:LINK
........................................................................... 786
...........................................................................................787
[SENSe:]PMETer<p>:MTIMe:AVERage:COUNt
.................................................................787
[SENSe:]PMETer<p>:MTIMe:AVERage[:STATe]
................................................................ 787
[SENSe:]PMETer<p>:ROFFset[:STATe]
............................................................................788
........................................................................................ 788
..........................................................................................788
[SENSe:]PMETer<p>:UPDate[:STATe]
..............................................................................789
..........................................................................................789
................................................................................ 789
CALibration:PMETer<p>:ZERO:AUTO ONCE
This command zeroes the power sensor.
Note that you have to disconnect the signals from the power sensor input before you
start to zero the power sensor. Otherwise, results are invalid.
Suffix:
<p>
.
Power sensor index
Example:
CAL:PMET2:ZERO:AUTO ONCE;*WAI
Starts zeroing the power sensor 2 and delays the execution of
further commands until zeroing is concluded.
Usage:
Event
Manual operation:
See
Setting Basic Spectrum RF Measurement Parameters