Remote Commands
R&S
®
FPL1000
862
User Manual 1178.3370.02 ─ 11
Suffix:
<n>
.
<m>
Parameters:
<Offset>
Numeric value
*RST:
0
Default unit: dB
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed[:STATe]
<State>
This command activates or deactivates a marker that defines a fixed reference point
for relative marker analysis.
If necessary, the command activates a marker and positions it on the peak power.
Subsequently, you can change the coordinates of the fixed reference independent of
the marker. The fixed reference is independent of the trace and is applied to all active
delta markers.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:DELT:FUNC:FIX ON
Switches on the measurement with fixed reference value for all
delta markers.
CALC:DELT:FUNC:FIX:RPO:X 128 MHZ
Sets the frequency reference to 128 MHz.
CALC:DELT:FUNC:FIX:RPO:Y 30 DBM
Sets the reference level to +30 dBm.
Manual operation:
See
10.8.8.8
Marker Peak Lists
Useful commands for peak lists described elsewhere
●
CALCulate<n>:MARKer<m>:PEXCursion
●
●
Chapter 10.8.8.3, "Configuring and Performing a Marker Search"