Network Operation and Remote Control
R&S
®
FPL1000
566
User Manual 1178.3370.02 ─ 11
–
The following instrument settings are changed:
Table 9-16: Instrument settings for emulation of 8566A/B, 8568A/B, 8594E; FSEA, FSEB, FSEM; FSEK
instruments
Model
# of Trace
Points
Start Freq.
Stop Freq.
Ref Level
Input Coupling
8566A/B
1001
2 GHz
22 GHz
0 dBm
AC
8568A/B
1001
0 Hz
1.5 GHz
0 dBm
AC
8560E
601
0 Hz
2.9 GHz
0 dBm
AC
8561E
601
0 Hz
6.5 GHz
0 dBm
AC
8562E
601
0 Hz
13.2 GHz
0 dBm
AC
8563E
601
0 Hz
26.5 GHz
0 dBm
AC
8564E
601
0 Hz
40 GHz
0 dBm
AC
8565E
601
0 Hz
50 GHz
0 dBm
AC
8594E
401
0 Hz
3 GHz
0 dBm
AC
Note
: The stop frequency indicated in the table may be limited to the corresponding
frequency of the R&S
FPL1000, if required.
9.4
Network and Remote Control Settings
Access
: [SETUP] > "N Remote"
The remote commands required to define these settings are described in
ter 10.10.6, "Configuring the Network and Remote Control"
Step-by-step instructions are provided in
Chapter 9.5, "How to Set Up a Network and
..................................................................................... 566
............................................................................................571
Returning to Manual Mode ("Local")
..................................................................... 576
9.4.1
General Network Settings
Access
: [SETUP] > "N Remote" > "Network" tab
The R&S
FPL1000 can be operated in a local area network (LAN), for example to con-
trol the instrument from a remote PC or use a network printer.
Network settings can only be edited in the firmware if a LAN cable is connected to the
R&S
FPL1000.
Network and Remote Control Settings