Remote Commands
R&S
®
FPL1000
649
User Manual 1178.3370.02 ─ 11
[SENSe:]ESPectrum<sb>:PRESet:STORe
<Standard>
This command saves the current SEM measurement configuration.
Standard definitions are stored in an xml file. The default directory for SEM standards
is
C:\Users\Public\Documents\Rohde-Schwarz\Analyzer\sem_std
.
Suffix:
<sb>
.
1 to 3
Sub block in a Multi-SEM measurement
Parameters:
<Standard>
String containing the file name.
You can save the file in a subdirectory of the directory men-
tioned above. In that case, you have to include the relative path
to the file.
Manual operation:
See
10.6.7.2
Controlling the Measurement
The following commands control the measurement itself.
INITiate<n>:ESPectrum
This command initiates a Spectrum Emission Mask measurement.
Suffix:
<n>
.
irrelevant
[SENSe:]SWEep:MODE
<Mode>
This command selects the spurious emission and spectrum emission mask measure-
ments.
You can select other measurements with
●
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>[:STATe]
Parameters:
<Mode>
LIST | AUTO | ESPectrum
AUTO
Turns on basic spectrum measurements.
ESPectrum
Turns on spectrum emission mask measurements.
LIST
Turns on spurious emission measurements.
*RST:
AUTO
Example:
SWE:MODE ESP
Configuring and Performing Measurements