Remote Commands
R&S
®
FPL1000
811
User Manual 1178.3370.02 ─ 11
POSitive
Positive peak
QPEak
Quasipeak (CISPR filter only)
SAMPle
First value detected per trace point
RMS
RMS value
AVERage
Average
CAVerage
CISPR Average (CISPR filter only)
CRMS
CISPR RMS (CISPR filter only)
*RST:
APEak
Example:
DET POS
Sets the detector to "positive peak".
Manual operation:
See
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO
<State>
This command couples and decouples the detector to the trace mode.
Suffix:
<n>
.
<t>
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
DET:AUTO OFF
The selection of the detector is not coupled to the trace mode.
Manual operation:
See
TRACe<n>:COPY
<TraceNumber>, <TraceNumber>
This command copies data from one trace to another.
Suffix:
<n>
.
Parameters:
<TraceNumber>
TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6
The first parameter is the destination trace, the second parame-
ter is the source.
(Note the 'e' in the parameter is required!)
Setting Basic Spectrum RF Measurement Parameters