Remote Commands
R&S
®
FPL1000
871
User Manual 1178.3370.02 ─ 11
Example:
CALC:MARK:FUNC:BPOW:AOFF
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:MODE
<Mode>
This command selects the way the results for a band power marker are displayed.
(Note: relative power results are only availabe for delta markers,
see .
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:MODE
Suffix:
<n>
.
<m>
Parameters:
<Mode>
POWer
Result is displayed as an absolute power. The power unit
depends on the
DENSity
Result is displayed as a density in dBm/Hz.
*RST:
POWer
Example:
CALC:MARK4:FUNC:BPOW:MODE DENS
Configures marker 4 to show the measurement results in
dBm/Hz.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:BPOWer:RESult?
This command queries the results of the band power measurement.
Suffix:
<n>
.
<m>
Return values:
<Power>
Signal power over the marker bandwidth.
Example:
Activate the band power marker:
CALC:MARK:FUNC:BPOW:STAT ON
Select the density mode for the result:
CALC:MARK:FUNC:BPOW:MODE DENS
Query the result:
CALC:MARK:FUNC:BPOW:RES?
Response:
20dBm/Hz
Usage:
Query only
Setting Basic Spectrum RF Measurement Parameters