![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 764](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987764.webp)
Remote Commands
R&S
®
FPL1000
764
User Manual 1178.3370.02 ─ 11
Table 10-5: Optimization parameters in zero span mode
Optimization mode
Description
DYNamic
The narrowest filter possible (depending on the RBW) is used.
SPEed
The widest filter possible (depending on the RBW) is used.
AUTO
A medium-sized prefilter is used.
Note: EMI measurements
For EMI measurements (using R&S
FPL1-K54), "Dynamic" mode is not supported.
"Auto" mode always uses "Speed" optimization.
Parameters:
<Mode>
*RST:
AUTO
Example:
SWE:OPT DYN
Selects optimization for dynamic range.
Manual operation:
See
[SENSe:]SWEep[:WINDow<n>]:POINts
<SweepPoints>
This command defines the number of sweep points to analyze after a sweep.
Note that the number of sweep points is limited to 10001 when measuring spurious
emissions.
Suffix:
<n>
.
Parameters:
<SweepPoints>
Range:
101 to 100001
*RST:
1001
Example:
SWE:POIN 251
Manual operation:
See
[SENSe:]SWEep:TIME
<Time>
This command defines the sweep time. It automatically decouples the time from any
other settings.
In the Spectrum application, the command decouples the sweep time from the span
and resolution and video bandwidths. Note that this command queries only the time
required to capture the data, not to process it. To obtain an estimation of the total cap-
ture and processing time, use the
command.
Parameters:
<Time>
refer to data sheet
*RST:
depends on current settings (determined automati-
cally)
Default unit: S
Setting Basic Spectrum RF Measurement Parameters