![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 267](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987267.webp)
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
267
User Manual 1178.3370.02 ─ 11
is indicated in the channel setup info, next to the RBW information. For details, see
"Resolution Bandwidth and Filter Types"
The RBW filter configured in the bandwidth settings is identical to the filter configured
in the EMI configuration.
Remote command:
[SENSe:]BANDwidth[:RESolution]:TYPE
RBW
Defines the resolution bandwidth. The available resolution bandwidths are specified in
the data sheet. Numeric input is always rounded to the nearest possible bandwidth.
If "Auto" is selected, the resolution bandwidth is coupled to the selected span (for span
> 0). If the span is changed, the resolution bandwidth is automatically adjusted.
If the resolution bandwidth is defined manually, a green bullet is displayed next to the
"RBW" display in the channel bar.
For more information see
Chapter 6.6.1.1, "Separating Signals by Selecting an Appro-
Note:
Restrictions.
●
For EMI measurements using the quasipeak detector, the 1
MHz RBW filter is not
"Resolution Bandwidth and Filter Types"
Remote command:
[SENSe:]BANDwidth[:RESolution]
[SENSe:]BANDwidth[:RESolution]:AUTO
Automatic Peak Search
If activated, a peak search is performed automatically for all active markers after each
sweep.
If Auto peak search
and
limit lines are active, the active markers are set to the peak
delta
values between the measured signal and the limit lines.
Note:
Auto Max Peak Search / Auto Min Peak Search
are not available for EMI measurements.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:PEAKsearch:AUTO
Dwell Time
Sets the dwell time for the EMI marker measurement.
For more information see
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DWELl
Final Test Detector
Defines the detector to be used for the final EMI test at the marker frequency.
This setting is only available if the EMI (R&S
FPL1-K54) measurement option is instal-
led. CISPR detectors are only available after selecting a CISPR RBW or filter (see
Measurements and Results