The Spectrum Application (RF Measurements)
R&S
®
FPL1000
162
User Manual 1178.3370.02 ─ 11
Label
Description
"Power Rel"
Peak power level within the range, relative to the "Tx Power"
"
Δ
Limit"
Deviation of the peak power level from the limit line
You can define in which detail the data is displayed in the result summary in the "List
Evaluation" settings (see
"List Evaluation (Results Configuration)"
default, one peak per range is displayed. However, you can change the settings to dis-
play only peaks that exceed a threshold ("Margin").
Detected peaks are not only listed in the Result Summary, they are also indicated by
colored squares in the diagram (optionally, see
settings).
Figure 6-26: Detected peak display in SEM measurement
Furthermore, you can export the results of the result summary to a file which can be
exported to another application for further analysis.
Results for SEM with multiple sub blocks
In the Spectrum application only, spectrum emissions can be measured for multiple
sub blocks of channels (see
"SEM with Multiple Sub Blocks ("Multi-SEM")"
on page 169 for details). Up to 3 sub blocks (with 2 gaps) can be defined. For each sub
block and each gap, the results described above are provided individually in the result
summary.
Measurements and Results