Remote Commands
R&S
®
FPL1000
906
User Manual 1178.3370.02 ─ 11
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:LIM:STAT ON
Switches on the limit check for limit line 1.
Manual operation:
See
CALCulate<n>:LIMit<li>:TRACe<t>:CHECk
<State>
This command turns the limit check for a specific trace on and off.
To query the limit check result, use
.
Suffix:
<n>
.
<li>
<t>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:LIM3:TRAC2:CHEC ON
Switches on the limit check for limit line 3 on trace 2.
Manual operation:
See
MMEMory:LOAD<n>:LIMit
<FileName>
Loads the limit line from the selected file in
.CSV
format.
Suffix:
<n>
.
irrelevant
Parameters:
<FileName>
String containing the path and name of the
CSV
import file.
Example:
MMEM:LOAD:LIM 'C:\TEST.CSV'
Manual operation:
See
MMEMory:STORe<n>:LIMit
<FileName>, <LimitLineName>
This command exports limit line data to an ASCII (CSV) file.
For details on the file format see
Chapter 6.11.2.4, "Reference: Limit Line File Format"
Setting Basic Spectrum RF Measurement Parameters