The Spectrum Application (RF Measurements)
R&S
®
FPL1000
313
User Manual 1178.3370.02 ─ 11
level. Relative values are indicated in percent for linear scaling, and in dB for logarith-
mic scaling.
6.5.2
Amplitude Settings
Access
: "Overview" > "Amplitude"
Amplitude settings determine how the R&S
FPL1000 must process or display the
expected input power levels.
The remote commands required to define these settings are described in
ter 10.8.3.1, "Amplitude Settings"
.......................................................................... 314
Setting the Reference Level Automatically (Auto Level)
...............................315
............................................................................................ 315
Reference Level
Defines the expected maximum input signal level. Signal levels above this value are
possibly not measured correctly, which is indicated by the "IF Overload" status display
("OVLD" for analog baseband or digital baseband input).
Defines the expected maximum reference level. Signal levels above this value are pos-
sibly not measured correctly. Signals above the reference level are indicated by an "IF
Overload" status display.
Amplitude and Vertical Axis Configuration