Remote Commands
R&S
®
FPL1000
680
User Manual 1178.3370.02 ─ 11
Example:
LIST:RANG2:DET AVER
Manual operation:
See
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STARt
<Frequency>
This command defines the start frequency of a spurious emission measurement range.
Make sure to set an appropriate span. If you set a span that is
●
smaller than the span the sweep list covers, the R&S
FPL1000 will not measure
the ranges that are outside the span - results may be invalid.
●
greater than the span the sweep list covers, the R&S
FPL1000 will adjust the start
frequency of the first range and the stop frequency of the last range to the span
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Frequency>
Numeric value.
*RST:
-12.75 MHz (range 1), -2.515 MHz (range 2), 2.515
MHz (range 3)
Default unit: Hz
Example:
LIST:RANG2:STAR 2MHZ
Manual operation:
See
[SENSe:]LIST:RANGe<ri>[:FREQuency]:STOP
<Frequency>
This command defines the stop frequency of a spurious emission measurement range.
Make sure to set an appropriate span. If you set a span that is
●
smaller than the span the sweep list covers, the R&S
FPL1000 will not measure
the ranges that are outside the span - results may be invalid.
●
greater than the span the sweep list covers, the R&S
FPL1000 will adjust the start
frequency of the first range and the stop frequency of the last range to the span
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Frequency>
Numeric value.
*RST:
-2.52 MHz (range 1), 2.52 MHz (range 2), 250.0
MHz (range 3)
Default unit: Hz
Example:
LIST:RANG2:STOP 5MHZ
Manual operation:
See
Configuring and Performing Measurements