![Ronde & Schwarz 1304.0004K03 Скачать руководство пользователя страница 820](http://html1.mh-extra.com/html/ronde-and-schwarz/1304-0004k03/1304-0004k03_user-manual_1484987820.webp)
Remote Commands
R&S
®
FPL1000
820
User Manual 1178.3370.02 ─ 11
Manual operation:
See
10.8.7.3
Using Trace Mathematics
The following commands control trace mathematics.
CALCulate<n>:MATH<t>[:EXPRession][:DEFine]
.............................................................. 820
......................................................................................820
...................................................................................821
...................................................................................... 821
CALCulate<n>:MATH<t>[:EXPRession][:DEFine]
<Expression>
This command selects the mathematical expression for trace mathematics.
Before you can use the command, you have to turn trace mathematics on.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Expression>
(TRACE1-TRACE2)
Subtracts trace 2 from trace 1.
(TRACE1-TRACE3)
Subtracts trace 3 from trace 1.
(TRACE1-TRACE4)
Subtracts trace 4 from trace 1.
(TRACE1-TRACE5)
Subtracts trace 5 from trace 1.
(TRACE1-TRACE6)
Subtracts trace 6 from trace 1.
Example:
CALC:MATH:STAT ON
Turns trace mathematics on.
CALC:MATH:EXPR:DEF (TRACE1-TRACE3)
Subtracts trace 3 from trace 1.
Manual operation:
See
CALCulate<n>:MATH<t>:MODE
<Mode>
This command selects the way the R&S
FPL1000 calculates trace mathematics.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Mode>
For more information on the way each mode works see
Setting Basic Spectrum RF Measurement Parameters