Remote Commands
R&S
®
FPL1000
621
User Manual 1178.3370.02 ─ 11
OBANdwidth | OBWidth
Occupied bandwidth measurement
CN
Carrier to noise ratio
CN0
Carrier to noise ration referenced to a 1
Hz bandwidth
Manual operation:
See
"Optimized Settings (Adjust Settings)"
See
See
[SENSe:]POWer:ACHannel:PRESet:RLEVel
This command determines the ideal reference level for the current measurement.
This automatic routine makes sure that the that the signal power level does not over-
load the R&S
FPL1000 or limit the dynamic range by too small a S/N ratio.
To determine the best reference level, the R&S
FPL1000 aborts current measurements
and performs a series of test sweeps. After it has finished the test, it continues with the
actual measurement.
To get a valid result, you have to perform a complete sweep with synchronization to the
sweep end. This is only possible in single sweep mode.
[SENSe:]POWer:TRACe
<TraceNumber>
This command selects the trace channel power measurements are performed on.
For the measurement to work, the corresponding trace has to be active.
Parameters:
<TraceNumber>
Range:
1 to 6
*RST:
1
Example:
POW:TRAC 2
Assigns the measurement to trace 2.
Manual operation:
See
10.6.3
Measuring Compression Points
All remote control commands specific to compression point measurements are descri-
bed here.
Configuring the Compression Point Measurement
............................................... 622
Programming Example: Measuring Compression Points
......................................625
Configuring and Performing Measurements